近赤外線法と水中体重法による体脂肪率の比較


File Description SizeFormat
iryou05_00_03_t.pdf510.46 kBAdobe PDFView/Open

Title: 近赤外線法と水中体重法による体脂肪率の比較
Other Titles: COMPARISON BETWEEN FITNESS ANALYZER AND THE UNDERWATER WEIGHING METHOD FOR ESTIMATION OF BODY COMPOSITION
Authors: 浦田, 秀子 / 大塚, 健作 / 西山, 久美子 / 勝野, 久美子 / 福山, 由美子 / 田原, 靖昭 / 綱分, 憲明
Authors (alternative): Urata, Hideko / Otsuka, Kensaku / Nishiyama, Kumiko / Katsuno, Kumiko / Fukuyama, Yumiko / Tahara, Yasuaki / Tsunawake, Noriaki
Issue Date: 31-Mar-1992
Citation: 長崎大学医療技術短期大学部紀要 = Bulletin of the School of Allied Medical Sciences, Nagasaki University. 1992, 5, p.15-22
Abstract: 近赤外線法(BFT-2000)による体脂肪測定器の有用性を検討するため,167名を対象に密度法である水中体重法と比較検討した.水中体重法とBFT-2000による体脂肪率の相関係数は0.888と有意な正の相関があったが,測定値の平均値はBFT-2000による場合が有意に高く,両法による測定値が一致するとは限らなかった.BFT-2000による測定値は水中体重法による体脂肪率が30%以下の群では高目に,30%以上では低目になり,両者による測定値の差(近赤外線法一水中体重法)と水中体重法による体脂肪率との間には負の相関があった. BFT-3000と水中体重法による測定値との間には相関係数は0.868で正の相関があり,BFT-2000より水中体重法に近い値であった. / The reliability of Fitness Analyzer BFT-2000, an instrument for estimation of body composition in human, was tested by comparing with the method of underwater weighing (UWW). For all 167 subjects, correlation coefficient of the values for %Fat estimated from the both methods was 0.888. However, the mean values for %FAT obtained from BFT-2000 and UWW were 21.0±7.6% and l7.4±8.6% respectively, and the former was significantly higher than the later (P< 0.01). By further analysis of the individual data, BFT-2000 had a tendency to overestimate as long as the values for %Fat from UWW is less than 30%, but to underestimate when the values from UWW is over 30%. BFT-3000, a new model of Fitness Analyzer was also tested. Correlation coefficients between values for %Fat estimated from BFT-3000, BFT-2000 and UWW were high (0.830~0.971). As compared with BFT-2000, the values from BFT-3000 were more close to the values from UWW.
Keywords: 近赤外線法 / FITNESS ANALYZER / 水中体重法 / 体脂肪率
URI: http://hdl.handle.net/10069/18151
ISSN: 09160841
Type: Departmental Bulletin Paper
Text Version: publisher
Appears in Collections:Volume 5